• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GEN II 7.5MM RESUR PAT 32MM; PRSTHSS,KNEE,PTLLOFMROTIBL,SMI-CNSTRND,CMNTD,PLYMER/METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GEN II 7.5MM RESUR PAT 32MM; PRSTHSS,KNEE,PTLLOFMROTIBL,SMI-CNSTRND,CMNTD,PLYMER/METAL/POLYMER Back to Search Results
Catalog Number 71932636
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code Injury (2348)
Date of Event 05/03/2016
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient presented excessive arthrofibrosis of right knee resulting in severe limited range of motion and development of scar tissue resulting in a closed manipulation.
 
Additional Manufacturer Narrative
The associated complaint devices were not returned.A clinical evaluation was conducted and it was reported that a revision surgery was performed due to excessive lack of motion, pain and infection.The exact cause of the reported infection could not be determined based solely on the information provided.It was further reported the patient presented arthrofibrosis of the right knee with severe limited range of motion and development of scar tissue resulting in a closed manipulation.Arthrofibrosis is a known possible complication after surgical intervention.Although the s&n devices cannot be excluded as to have caused or contributed to the reported event.The explanted device was destroyed.Therefore, based on insufficient information, a thorough clinical assessment cannot be performed at this time and the patient impact beyond the reported arthrofibrosis cannot be determined.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GEN II 7.5MM RESUR PAT 32MM
Common Device Name
PRSTHSS,KNEE,PTLLOFMROTIBL,SMI-CNSTRND,CMNTD,PLYMER/METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8525396
Report Number1020279-2019-01602
Device Sequence Number16843679
Product Code JWH
UDI-Device Identifier00885556056721
UDI-Public00885556056721
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K951987
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/26/2026
Device Catalogue Number71932636
Device Lot Number16AM19244
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/18/2019
Supplement Date Received by Manufacturer03/18/2019
Initial Report FDA Received Date04/17/2019
Supplement Report FDA Received Date10/29/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
-
-