| Catalog Number |
103453 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Cardiac Arrest (1762)
|
| Date of Event |
03/25/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient experienced a cardiac arrest while undergoing treatment with a phoenix dialysis machine.The cause of the event was not reported.Treatment was not reported.At the time of this report the patient outcome was not reported.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was evaluated on site.The phoenix device received functional testing.The device passed functional testing and no issues were found.No previous alarms were found.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.The reported issue was not verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this serial number.A device history review revealed no issues that could have caused or contributed to the reported issue.No device failure or malfunction was identified that could have caused or contributed to the reported event; therefore, the phoenix device is no longer considered suspect product in this event.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|