The patient was undergoing a coil embolization procedure in the internal iliac artery (iia) using a lantern delivery microcatheter (lantern).During the procedure, the lantern was flushed and advanced to the target vessel.The physician then found that the lantern had become flattened and the guidewire was not able to advance through.The lantern was reported to be collapsed at approximately the middle to proximal end, and was therefore removed from the procedure.The procedure was completed using a new lantern.There was no report of an adverse effect to the patient.
|
Results: the lantern was ovalized and had multiple consecutive kinks from approximately 144.5 ¿ 149.5 cm from the hub.During functional testing, resistance was experienced at the ovalization while advancing a 0.025¿ stainless steel mandrel through the lantern and the mandrel could not advance any further.Conclusions: evaluation of the returned lantern revealed distal ovalizations and consecutive kinks.If the peel away sheath included with the lantern is not properly utilized when advancing the lantern into parent device, damage such as distal ovalizations and consecutive kinks may occur.This damage likely contributed to the reported complaint.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|