| Model Number |
N/A |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Codes |
Extraskeletal ossification (1428); Erosion (1750); Hemorrhage/Blood Loss/Bleeding (1888); Low Blood Pressure/ Hypotension (1914); Tachycardia (2095)
|
| Date of Event |
02/22/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Zimmer biomet complaint (b)(4).The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Report sources: brice, h., lacroix, v., pirotte, t.& docquier, p.(2018)."lung middle lobe laceration needing lobectomy as complication of nuss bar removal." case reports in orthopedics, volume 2018, pages 1-6.Https://doi.Org/10.1155/2018/8965641.Notrica, d., mcmahon, l., johnson, k., velez, d.A., mcgill, l.C., & jaroszewski, d.E.(2014)."life-threatening hemorrhage during removal of a nuss bar associated with sternal erosion." annals of thoracic surgery, vol.98 (no.3), pages 1104¿1106.Http://dx.Doi.Org/10.1016/j.Athoracsur.2013.10.097.
|
| |
|
Event or Problem Description
|
|
It was reported in a journal article that the patient experienced life threatening bleeding after removal of a pectus bar.A (b)(6) woman presented for elective removal of her nuss pectus bars.She had undergone repair of pectus excavatum 3 years earlier.At her 3-year followup, she had excellent correction with resolution of all preoperative symptoms.During the bar removal, the prior incisions were opened bilaterally.Both bars had moderate heterotopic ossification as well as fibrin sheath encasement.Both bars were freed of all bony fragments and completely straightened.The lower bar was removed first without difficulty.The upper bar initially showed significant resistance to removal.With an additional effort, the straightened bar was removed.Immediately after the bar¿s exit, a rapid flow of non-pulsatile bright red blood flowed from both incisions.The patient rapidly became tachycardiac and hypotensive.Transfusion protocols and conversion to anterior thoracotomy were initiated.The right inframammary incision was extended to open the right chest.There was no blood in the right chest or pericardium.The anterior thoracotomy incision was extended to the left side.Bone cutters were required to cut across the heterotopic ossification where the bars crossed the ribs.The hemorrhage stopped spontaneously during exploration after extension of the left thoracotomy.Bleeding cessation correlated temporally with hypotension and the administration of epinephrine.At the site of bar erosion into the sternum, the internal mammary arteries (imas) were not visible and were assumed to be incorporated into the sternum.As this was the suspected site of bleeding, proximal ligation was performed bilaterally of the imas to decrease the risk of rebleeding.Chest drains were placed.The patient did not experience cardiac arrest or require chest compressions.Estimated blood loss was approximately 3.5 l.She received 4 units of packed red blood cells, 1 unit of fresh-frozen plasma, and 1 pheresis of platelets.At discharge, her hemoglobin was 8.4 mg/dl.Her repair has remained stable, and she recovered fully.No additional patient consequences were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The surgeon sent a response stating that the bleeding the patient experienced was related to patient condition.
|
| |
|
Event or Problem Description
|
|
The surgeon sent a response stating that the bleeding the patient experienced was related to patient condition.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Based on the information and materials provided, the complaint is confirmed.Functional testing and inspections could not be performed due to the parts not being returned.Device history record (dhr) review was unable to be performed as the lot number of the device involved in the event is unknown.In a response from the surgeon, the surgeon stated the bleeding experienced was due to patient condition and was not related to the implant.There are no indications of manufacturing defects.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
This follow-up report is being submitted to relay additional information.
|
| |
|