| Model Number |
0602870 |
| Medical Device Problem Codes |
Component Incompatible (1108); Failure to Infuse (2340)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device history records have been reviewed with special attention to the raw materials, subassemblies, manufacturing process and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.The device has been returned for evaluation; the investigation is confirmed for the alleged failure to infuse.Based upon the available information, the definitive root cause is unknown.The device is labeled for single use.
|
| |
|
Event or Problem Description
|
|
This report summarizes one (1) malfunction event.A review of the event indicated that model 0602870 implanted port experienced a component incompatibility and a failure to infuse.This event was reported from a single source.This event involved a patient with no reported injury.The patient was reported as a female weighing (b)(6) kgs whose age was not provided.
|
| |
|