• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MMJ SA DE CV (USD) INVOS; OXIMETER, TISSUE SATURATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MMJ SA DE CV (USD) INVOS; OXIMETER, TISSUE SATURATION Back to Search Results
Model Number PMSENS71-A-20
Medical Device Problem Code Low Readings (2460)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/19/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during procedure, the device showed "signal error detected" and the rso2 values the lower or were not shown.It was indicated that there was no patient injury.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Evaluation summary: one device was received for evaluation.The reported event was not confirmed.The device passed the all test criteria.Unit gave consistent readings with no errors.Inspection of the sensor found no anomalies.The investigation found the device to function normally.No new formal investigation is required, the event will be included in trending and monitoring.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INVOS
Common Device Name
OXIMETER, TISSUE SATURATION
Manufacturer (Section D)
MMJ SA DE CV (USD)
ave henequen no 1181 desarroll
ciudad juarez 32590
MX  32590
MDR Report Key8529804
Report Number2936999-2019-00293
Device Sequence Number1840015
Product Code MUD
UDI-Device Identifier20884521556970
UDI-Public20884521556970
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K182868
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPMSENS71-A-20
Device Catalogue NumberPMSENS71-A-20
Device Lot NumberAD182501
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/02/2019
Initial Date Received by Manufacturer 04/16/2019
Supplement Date Received by Manufacturer10/04/2019
12/05/2019
Initial Report FDA Received Date04/18/2019
Supplement Report FDA Received Date10/31/2019
12/18/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age75 YR
Patient Weight82
-
-