MAUDE Adverse Event Report: BIOFIRE DIAGNOSTICS, LLC FILMARRAY ® BLOOD CULTURE IDENTIFICATION (BCID) PANEL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOFIRE DIAGNOSTICS, LLC FILMARRAY ® BLOOD CULTURE IDENTIFICATION (BCID) PANEL Back to Search Results
Model Number RFIT-ASY-0126
Medical Device Problem Code False Positive Result (1227)
Health Effect - Clinical Codes Death (1802); Renal Disease, End Stage (2039)
Date of Event 03/19/2019
Type of Reportable Event Death
Additional Manufacturer Narrative
Investigation: (b)(6) reported a discrepant (b)(6) result on the filmarray bcid panel after testing a blood sample from a (b)(6) year old, male patient.The patient did not have a prior (b)(6) infection.The patient was on ciprofloxacin and rifaximin for spontaneous bacterial peritonitis (sbp) prophylaxis and hepatic encephalopathy, respectively, at the time of the blood draw.Two sets of blood culture bottles were collected on (b)(6) 2019, and one set flagged (b)(6).On (b)(6) 2019, a (b)(6) blood culture (bactalert sn anaerobic bottle, lot# 1052516, exp.November 20, 2019) was tested on the filmarray bcid panel and the result was (b)(6).The customer used id by staph latex, and the customer stated culture results did not show signs of a mixed culture.Four isolates were also tested this same day for susceptibility by microscan (which shows (b)(6) sensitivity) and the preliminary result was (b)(6).On (b)(6) 2019, the patient was put on (b)(6).The customer reported that the (b)(6) treatment was partly due to the filmarray bcid result, and partly due to a possible beta-lactam allergy.After one dose of (b)(6), the patient developed tongue swelling and was switched to (b)(6).On (b)(6) 2019, the same blood culture was repeated on the filmarray bcid panel and the result was (b)(6).The patient developed acute kidney injury (scr 1.21 mg/dl) on (b)(6) 2019, which according to the customer was "likely multifactorial" as the patient had hypovolemia, hepatorenal syndrome, and was on nephrotoxic medications (contrast on (b)(6) 2019, and (b)(6) on (b)(6) 2019).The patient's serum creatinine continued to increase the following three days (1.40, 1.69, 1.81 mg/dl respectively) and was reported as 1.62 mg/dl when the patient was admitted to hospice on (b)(6) 2019.Antibiotic treatment was discontinued when the patient was admitted to hospice care and the patient passed away on (b)(6) 2019.The customer reported that they followed decontamination protocol.Please note: there is a reasonable risk that (b)(6) may have contributed to increase kidney dysfunction.However, medications prior to the introduction of (b)(6) (ciprofloxacin, rifaxamin and contrast dye) can all contribute to renal insufficiency.Co-administration of (b)(6) with other nephrotoxic agents (in this case (b)(6)) has been associated with greater severity of aki and a lower probability of renal recovery.Incomplete renal recovery accelerates the progression of chronic kidney disease (ckd), and may result in higher rates of cardiovascular morbidity and fatality.At the time of (b)(6) administration, the serum creatinine levels were at the high end of the normal range (review of vancomycin-induced renal toxicity: an update.Ther adv endocrinol metab 2016 jun; 7(3): 136-147).The patient had other serious comorbidities that most likely contributed to the mortality including hypovolemia and hepatorenal syndrome which contributed to a high risk for sbp (hence ciprofloxacin prophylaxis), and hepatic encephalopathy (rifaxamin).However based on the information submitted it cannot be determined if the (b)(6) was responsible for the aki or increased the severity of the aki since the patient was initially on two drugs and then was co-administered with a drug that all could have resulted in or increased the risk for and the severity of kidney injury.Quality control (qc) records for pouch lot# 455919 (kit lot# 0116219) were reviewed.This pouch lost passed the qc criteria and was found within specifications.No run malfunction was observed and the filmarray instrument (serial# (b)(4)) was working within design specifications.Conclusion: the investigation concluded that the most likely cause for the discrepant result was a heterogeneous mixture of (b)(6) resistance as well as sensitivity/specificity differences between the filmarray bcid panel, standard culture, and traditional ast methods.The filmarray bcid panel does not specifically attribute meca-mediated methicillin resistance to either s.Aureus or other staphylococcus species.Unlike ast methods, filmarray bcid reports detections from the blood culture population as a whole and not from a single/few isolated colonies.The multi-species assays detect the most prevalent coagulase-negative staphylococcus (cons) species encountered in blood culture specimens and can also react with high levels of s.Aureus.No (b)(6) species were observed in culture which may indicate a mixed population of (b)(6).However, this is a large and diverse group and detection by filmarray bcid panel assays is variable (filmarray bcid panel instruction booklet, table 7, "multiple assay interpretations" section, https://www.Online-ifu.Com/iti0045).It is also possible that the (b)(6) contained (b)(6) but this mechanism was not expressed or inactivated.It is estimated that a small percentage (approximately 3% but likely higher) of s.Aureus strains that are positive for meca via pcr are highly susceptible to methicillin-like antibiotics as reported by ast methods [1-4].In addition, accurate identification of mrsa from mssa is further complicated by phenotypic variability in methicillin susceptibility, also known as heteroresistance [4-5].Other studies also report cases of s.Aureus strains reverting from methicillin susceptible to methicillin resistant during antibiotic treatment [4-5].Detection of bcid targets does not imply that the corresponding organisms are infectious or the causative agents for clinical symptoms.Results from the bcid panel must be correlated with the clinical history, epidemiological data, and other data available to the clinician evaluating the patient.Late signatures of staphlyococcus1 and staphylococcus2 suggest the meca detection could be attributed to a cons species, which potentially could be a skin flora contaminant.During panel development, meca (staphylococcus detected, s.Aureus detected) showed robust performance comparing to phenotypic antimicrobial susceptibility testing (ast) methods, with a sensitivity of 97.7% (95% ci 93.4 - 99.5%) and specificity of 96.7% (95% ci 91.7 - 99.1%) for prospective staphylococcus, s.Aureus detected (filmarray bcid panel instruction booklet, table 26, https://www.Online-ifu.Com/iti0045).References: hososaka y, hanaki h, endo h et al.2007.Characterization of oxacillin-susceptible meca-positive staphylococcus aureus: a new type of mrsa.J infect chemother; 13:79-86.Kumar va, steffy k, chatterjee m et al.2013.Detection of oxacillin-susceptible meca-positive staphylococcus aureus isolates by use of chromogenic medium mrsa id.J clin microbiol.51:318-9.Salem-bekhit mm.2014.Phenotypic and genotypic characterization of nosocomial isolates of staphylococcus aureus with reference to methicillin resistance.Trop j pharm res; 13:1239-46.Proulx mk, et al.2015.Reversion from methicillin susceptibility to methicillin resistance in staphylococcus aureus during treatment of bacteremia.J infect dis.213(6):1041-48.Sakoulas g, et al.2001.Methicillin-resistant staphylococcus aureus: comparison of susceptibility testing methods and analysis of meca-positive susceptible strains.J clin microbiol.39(1):3946-51.
 
Event or Problem Description
(b)(6) reported a discrepant (b)(6) result on the filmarray blood culture identification (bcid) panel after testing a blood sample from a (b)(6) year old, male patient.Due to the filmarray bcid panel result, the customer reported that treatment was altered and there was potential harm to the patient.Ultimately, the patient developed acute kidney injury (aki) and expired on (b)(6) 2019.Upon investigation, no malfunction occurred and the system was working within specification.The investigation concluded that the most likely cause for this discrepant result was sensitivity/specificity differences between the filmarray bcid panel, standard culture, and ast methods.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FILMARRAY ® BLOOD CULTURE IDENTIFICATION (BCID) PANEL
Common Device Name
FILMARRAY ® BLOOD CULTURE IDENTIFICATION (BCID) PANEL
Manufacturer (Section D)
BIOFIRE DIAGNOSTICS, LLC
515 colorow drive
salt lake city UT 84108
MDR Report Key8529831
Report Number3002773840-2019-00003
Device Sequence Number11244917
Product Code PEN
UDI-Device Identifier00815381020086
UDI-Public00815381020086
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K181493
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/30/2020
Device Model NumberRFIT-ASY-0126
Device Catalogue NumberRFIT-ASY-0126
Device Lot Number0116219
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/25/2019
Initial Report FDA Received Date04/18/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient Age66 YR
-
-