Device Problem Moisture Damage (1405)
Patient Problems Oversedation (1990); No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Device evaluation: the pump has been returned and evaluated by product analysis on 17-apr-2019.With the following findings: evaluation revealed contamination under all of the keypad button contacts.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.(b)(6).
 
Event Description
The pump was returned for investigation.Evaluation revealed contamination under all of the keypad button contacts.There was no adverse event associated with this complaint.This report is made based on results of investigation completed on 17-apr-2019.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Type of Device
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS LLC
965 chesterbrook blvd
wayne PA 19087
Manufacturer (Section G)
ANIMAS LLC
965 chesterbrook blvd
wayne PA 19087
Manufacturer Contact
kristen lopolito
965 chesterbrook blvd
wayne, PA 19087
MDR Report Key8530121
MDR Text Key142535262
Report Number2531779-2019-02715
Device Sequence Number1
Product Code OYC
Combination Product (y/n)Y
PMA/PMN Number
P130007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial
Report Date 04/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/18/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/14/2019
Was the Report Sent to FDA? Yes
Date Manufacturer Received04/17/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age44 YR
-
-