Model Number PCB00
Medical Device Problem Codes Break (1069); Contamination (1120)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/21/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If explanted, give date: not applicable as to date the lens remains implanted and therefore not explanted.(b)(4).The intraocular lens (iol) is not returning for evaluation as it remains implanted; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, labeling, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that during implantation into the patient's right eye (od), white sediments were observed around the intraocular lens (model pcb00) and scratches.Reportedly, the surgeon prepared the lens himself with sufficient viscoelastic in the injector, he filled the chamber with viscoelastic as he always does, no resistance was noticed and nothing to raise concern for.However, after the injection into the eye, whitish elements floating around the lens were noticed.It looked like the lens had rubbed against the cartridge, it looked like shavings.The lens remains implanted.The account commented that the preloaded delivery system is not available for evaluation.The same issue was reported on three (3) pcb00s.On one of them, the surgeon noticed quite a bad scratch on the lens optic, although not on the visual axis.This reported is the second of the three mdr.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS ITEC PRELOADED 1-PIECE IOL
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key8530609
Report Number2648035-2019-00476
Device Sequence Number1732065
Product Code HQL
UDI-Device Identifier05050474558403
UDI-Public(01)05050474558403(17)190906
Combination Product (Y/N)Y
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/11/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/06/2019
Device Model NumberPCB00
Device Catalogue NumberPCB0000270
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 03/27/2019
Supplement Date Received by Manufacturer10/25/2020
Initial Report FDA Received Date04/19/2019
Supplement Report FDA Received Date11/12/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-