ETHICON ENDO-SURGERY, LLC. LIGACLIP*SINGLE CLIP APPLIER; APPLIER, SURGICAL, CLIP Back to Search Results
Catalog Number LC410
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Only event year known: 2019.Device analysis: the analysis results found that the lc410 device was received with the handle coating damaged, otherwise with no apparent damaged.The device was tested for functionality.Upon functional testing of the device, the instrument loaded, retained and deployed 6 clips as intended.The instrument was fully functional and conforming to our manufacturing requirements.No conclusion could be reached as to what may have caused the found condition of the handle coating.Please note that as stated in the ifu: "all surgical instruments are subject to a degree of wear and tear because of normal use.A regular and precise visual check of the instrument should be made before each use.The batch history records were reviewed and certed by external manufacturing that the manufacturing criteria was met prior to the release of the equipment.The certificate records are accessible through external manufacturing.
 
Event or Problem Description
It was reported that the jaws of the clip appliers are misaligned, resulting in the clips falling out of the jaws of the instrument during use or not locking on tissue during use.No patient involvement was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
APPLIER, SURGICAL, CLIP
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
MDR Report Key8531294
Report Number3005075853-2019-18177
Device Sequence Number1515472
Product Code GDO
UDI-Device Identifier10705036012702
UDI-Public10705036012702
Combination Product (Y/N)N
Initial Reporter StateND
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberLC410
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/10/2019
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/25/2019
Initial Report FDA Received Date04/19/2019
Patient Sequence Number1
-
-