Model Number 9004
Medical Device Problem Codes Retraction Problem (1536); Material Rupture (1546); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 03/21/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Analysis/device evaluation: upon receipt, visual examination of the catheter revealed multiple kinks along the length of the catheter.Further inspection noted the distal end of the balloon portion was returned with the device.Examination of the balloon material noted it was torn with jagged edges and surface abrasions at the initiation of the torn balloon material.The inner lumen is severely stretched and the proximal marker band is displaced and damaged in a slight oval shape.The distal end of the returned inner lumen is torn and the edges are jagged.There is shearing damage, proximal to the distal end of the inner lumen.Due to the returned condition of the device performance testing was unable to be performed.A lot history review revealed this is the only complaint associated with allegations of material rupture with retraction issues resulting in device separation for this lot.A device history record (dhr) review was performed and noted the device was manufactured to specification prior to being released for distribution.Conclusion: returned product analysis confirmed the reported material rupture and difficulty re-sheathing the balloon.Based on the information provided and analysis results, the material rupture was likely caused by treating a calcified non-target lesion in the common iliac artery.The force applied to the catheter during removal though the introducer sheath likely contributed to the balloon material separating from the catheter.There was nothing found to indicate there was a design, manufacturing or component anomaly which would have caused or contributed to this event.If additional information becomes available, a supplement report will be submitted with all relevant information.
 
Event or Problem Description
It was reported a lutonix percutaneous transluminal drug coated balloon (dcb) dilatation catheter allegedly ruptured on its second inflation at 14 atmospheres (atms) while treating a non-target lesion in the common iliac.The health care professional (hcp) used a contralateral approach with a 7 french ansel introducer sheath over an unknown 035 guidewire to treat the target lesion in the superficial femoral artery (sfa).The hcp reportedly used a hawk one atherectomy over an 014 guidewire to prepare the target lesion in the sfa.Reportedly, the 014 guidewire was exchanged for an 035 guidewire in preparation to advance the lutonix 035 catheter.The hcp then reached the target lesion with the lutonix dcb and allegedly inflated the device to 8 atms.After successful treatment of the sfa, the hcp pulled the lutonix dcb to a stenosed region of the common iliac artery and inflated the lutonix dcb to 14 atms.Reportedly, the lutonix dcb ruptured during this inflation attempt.Upon attempts to remove the device, reportedly the balloon would not re-sheath, resulting in the balloon portion becoming stuck in the introducer sheath.Eventually, the hcp tore the balloon and the balloon portion became lodged in the common iliac artery over the aortic bifurcation.The hcp obtained an additional access from the opposite side of the original patient access.The balloon portion was reportedly fixated to the artery wall with two kissing stents which completed the procedure.The lutonix dcb was returned for evaluation.No further adverse patient outcomes were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUTONIX 035 DRUG COATED BALLOON PTA CATHETER
Common Device Name
DRUG COATED BALLOON PTA CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (GFO)
289 bay road
queensbury NY 12804
MDR Report Key8531615
Report Number3006513822-2019-00060
Device Sequence Number1839803
Product Code ONU
UDI-Device Identifier00801741123245
UDI-Public(01)00801741123245
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P130024
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/15/2020
Device Model Number9004
Device Catalogue NumberLX3513051505F
Device Lot NumberGFBY1837
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/09/2019
Event Location Hospital
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/22/2019
Initial Report FDA Received Date04/19/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age86 YR
Patient Weight71
-
-