| Model Number |
6DCT |
| Medical Device Problem Codes |
Leak/Splash (1354); Use of Device Problem (1670)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Information (3190)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, the device had been used for three weeks and noticed that cuff slowly deflates but not to the point that the ventilator would alarm.The device was inflated with 8-9ml of air.The patient has not been re-cannulated due to the issue.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: one sample of device was received for evaluation.An inflation/deflation test was performed and it was observed the cuff held the air.Additionally the sample was submerged into a container filled with water and no leaks were noticed.No reported event was observed in the received sample.Information has been added to the database and trends will continue to be monitored.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, the device had been used for three weeks and noticed that cuff slowly deflates but not to the point that the ventilator would alarm.The device was inflated with 8-9ml of air.The patient has not been de-cannulated and there was no harmnoted.The cuff constantly re-inflated to resolved the issue.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|