CONCORD MANUFACTURING LIBERTY SELECT CYCLER ASSY(NON-VALUATED); SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
|
Back to Search Results |
|
Catalog Number RTLR180343 |
Device Problems
Thermal Decomposition of Device (1071); Melted (1385)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/21/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Plant investigation: the actual device was returned to the manufacturer for physical evaluation.An exterior visual inspection of the returned cycler showed no signs of physical damage.Upon power up, the cycler touch screen test failed.When powering on the cycler, the ok, stop and up/down arrow push buttons illuminated; however, the front panel touch screen remained blank.It was identified that the cause for the blank screen was due to a shorted transformer (t1) on the inverter board.The inverter board is located on the rear of the touch screen.A known good inverter board was installed and the display became fully operational.The plastic front panel assembly ground was melted and split apart due to the heat generated from contact with the inverter board.Three of the pump assembly screws that anchor the pump assembly to the cycler plate were missing.An internal visual inspection of the returned cycler encountered no other discrepancies.A review of the device manufacturing records was conducted by the manufacturer.There were no deviations or non-conformances during the manufacturing process.In addition, a device history record (dhr) review was performed and verified that the results of the in-progress and final quality control (qc) testing met all requirements.Upon completion of the evaluation, the reported issue was confirmed and the cause was determined to be a burnt transformer on the inverter board.The cycler was refurbished following the evaluation.
|
|
Event Description
|
It was reported that the screen of a patient¿s liberty select cycler went blank during drain 0 of their peritoneal dialysis (pd) treatment.The power cord was properly connected in both ends and the cycler was plugged into a working outlet.The ok and stop keys were on.The cycler was rebooted; however, the screen remained blank.At that point in time, the technical support representative advised the patient to discontinue use of the cycler and to notify their peritoneal dialysis registered nurse (pdrn) of the event.A replacement cycler was issued to the patient.It was reported that an alternate treatment option was available.Upon follow up, the pdrn confirmed that there were no adverse events or medical intervention required as a result of the reported event.The patient was able to complete treatment with manual supplies.The cycler was returned to the manufacturer and a replacement cycler was provided and received.Upon physical evaluation of the cycler by the manufacturer, it was identified that the transformer on the inverter board was burned.
|
|
Search Alerts/Recalls
|
|
|