Catalog Number 72200616SR |
Device Problems
Device Remains Activated (1525); Physical Resistance/Sticking (4012)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Event Description
|
It was reported that, durin an unknown procedure, the motor drive unit hand control oscillating button was stuck; it was slow to release causing it to continually run and/or not oscillate.The device was connected to the control unit.There was not any error message displayed.The procedure was successfully completed with the same device and using a foot pedal.There was not surgical delay as consequence.The patient did not suffer any injury.
|
|
Manufacturer Narrative
|
The device was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of oscillate button running continuously could not be reproduced.Product passed functional and high speed testing (100-10,000 rpms) for 5 minutes each in forward, reverse and oscillate directions.Unit passed functional tests on the dyonics power, dii and dii eip test control units with and without footswitch.Oscillate button does not stick and mdu did not run continuously during functional testing.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
|
|