• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS PONSKY NON-BALLOON REPLACEMENT GASTROSTOMY TUBE; FEEDING DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS PONSKY NON-BALLOON REPLACEMENT GASTROSTOMY TUBE; FEEDING DEVICE Back to Search Results
Model Number 000702
Medical Device Problem Code Material Perforation (2205)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Based upon the severity level and lot quantity, a dhr review is not required; however, the dhr was requested to review manufacturing records.Manufacturing records were reviewed and there were not found evidence that the failure mode reported in this complaint is caused by the mfg.Process.One ponsky non balloon replacement gastrostomy tube was returned for evaluation.Visual and microscopic evaluations were performed.The investigation is confirmed for hole in the feeding dome, as a small split was observed on the bumper.The split was observed on the opposite side of the obturator insertion point.The definitive root cause could not be determined based upon available information.The device is labeled for single use.
 
Event or Problem Description
This report summarizes one malfunction event.A review of this event indicated that model 000702 feeding device experienced material perforation.This report was received from one source.There was one patient involved with no patient consequences.Patient age, weight, and gender were not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PONSKY NON-BALLOON REPLACEMENT GASTROSTOMY TUBE
Common Device Name
FEEDING DEVICE
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key8532541
Report Number3006260740-2019-01061
Device Sequence Number16624483
Product Code KNT
UDI-Device Identifier00801741036293
UDI-Public(01)00801741036293
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number000702
Device Catalogue Number000702
Device Lot NumberRECS2016
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/31/2019
Initial Report FDA Received Date04/19/2019
Patient Sequence Number1
-
-