MAUDE Adverse Event Report: BAXTER HEALTHCARE - MOUNTAIN HOME HOMECHOICE AUTOMATED PD SET WITH CASSETTE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - MOUNTAIN HOME HOMECHOICE AUTOMATED PD SET WITH CASSETTE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number 5C4469C
Device Problems Disconnection (1171); Fluid/Blood Leak (1250)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/11/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(6).The device was received and is currently awaiting evaluation.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a homechoice cassette leaked at the patient connector heat seal bead.This occurred during patient use for peritoneal dialysis therapy.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Additional information : the device was received for evaluation.A visual inspection was performed with the naked eye and noted a leak at the patient connector heat seal bead.Functional testing including clear passage testing and clamp function testing were performed with no issues noted related to the reported event.An additional defect was identified during the clamp function testing that was not related to the reported event.Leak testing failed due to the leak at the patient connector heat seal bead.The direct cause of the event was determined to be manufacturing related.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.A nonconformance has been opened to address this issue.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - MOUNTAIN HOME
mountain home AR
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key8533335
MDR Text Key142551169
Report Number1416980-2019-02113
Device Sequence Number1
Product Code FKX
UDI-Device Identifier00085412087375
UDI-Public(01)00085412087375
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K923065
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date08/29/2023
Device Catalogue Number5C4469C
Device Lot NumberH18H29101
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/04/2019
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/26/2019
Initial Date FDA Received04/19/2019
Supplement Dates Manufacturer Received06/04/2019
Supplement Dates FDA Received07/01/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-