SMITH & NEPHEW, INC. GII PLASTIC REAMTHRU FEM TRL SZ 7 RT; PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY
|
Back to Search Results |
|
| Catalog Number |
71436397 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/01/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during tkr the femoral component trial broke on impaction.Delay of less than 30min reported.No injury reported.A back up device was available.
|
| |
|
Additional Manufacturer Narrative
|
|
The associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.One of the condyles fractured off the device at the threaded insert assembly point.The threaded insert was not returned.The device was manufactured in 2017 and exhibits signs of moderate wear/usage.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
|
| |
|