| Model Number |
MMT-523NAP |
| Medical Device Problem Code |
Device Difficult to Program or Calibrate (1496)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/11/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that they received button error alarm.The customer's blood glucose level was 181 mg/dl at the time of incident.Troubleshooting was performed and customer stated that the buttons are not responding.Advised the insulin pump will need to be replaced and to discontinue use of the pump and revert to backup plan.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with no (act, up and escape) button response due to corroded keypad traces.No button error alarm during testing noted.Unable to perform all functional testing including the displacement, operating currents, a21 error test, rewind, basic occlusion, occlusion, prime a33 and excessive no delivery test due to buttons not responding.
|
| |
|