| Catalog Number |
75447540 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
03/31/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Transpedicular screw levels implanted: l4-l5 it was reported that, post-op, the implanted screw was fatigued.The physician tried to remove stem from the screw but it was not allowed.The body of the screw remained in the patient."alif souvering" was performed as a result of alleged event.Patient did not achieve solid fusion.No other info was available.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: product analysis: visual and optical examination of mas bone screw confirm breakage at the base (neck) of the bone screw head and thread damage possible from the removal process.Microscopic examination of the surface reveals significant secondary (post-fracture) damage to the fracture surface.Undamaged areas of fracture surface revealed gently convex striations through the cross-sectional area of the bone screw neck.This type of damage is consistent with cyclic fatigue followed by overload once the fatigue had weakened the material.Additional information: d10, h3, h6 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|