• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AUTOSOFT 90; INSET II 60/6 GREY TCAP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AUTOSOFT 90; INSET II 60/6 GREY TCAP Back to Search Results
Lot Number 5231109
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code Elevated ketones/Diabetic Ketoacidosis (2364)
Type of Reportable Event Serious Injury
Event or Problem Description
Unomedical reference number (b)(4).Event occurred in the united states.The patient reported that cannula was bent at the infusion site, preventing delivery of insulin which laid to high blood glucose level (498 mg/dl).The patient requested bolus on the pump.The patient was hospitalized for high blood glucose level and received fluids through intravenous route and zolfran which resolved the issue.The infusion set was in use for 2 days.The patient was hospitalized for 14 days.Unomedical do not see bent/kinking as being related to human factors, but rather as a training issue including correct choices of insertion sites and infusion sets and cannula length.Furthermore, the soft cannula is a flexible material that during use and upon removal can bend slightly.No further information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOSOFT 90
Common Device Name
INSET II 60/6 GREY TCAP
MDR Report Key8535281
Report Number3003442380-2019-01139
Device Sequence Number399797
Product Code FPA
UDI-Device Identifier05705244014305
UDI-Public05705244014305
Combination Product (Y/N)N
PMA/510(K) Number
K032854
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device No Information
Device Expiration Date03/01/2021
Device Lot Number5231109
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/02/2018
Initial Report FDA Received Date04/22/2019
Patient Sequence Number1
-
-