| Model Number |
80618 |
| Medical Device Problem Codes |
Failure to Advance (2524); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/07/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: investigation completed on the returned device.Visual analysis revealed a fracture located 34.7cm from the hub.The device was not completely separated, as the inner liner was still attached.The fracture looked consistent with a bending force break.A.014 guidewire was inserted into the device but would not pass the kinked/fractured area of the shaft.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.
|
| |
|
Event or Problem Description
|
|
Reportable based on analysis completed (b)(6) 2019.It was reported that the catheter was kinked.A direxion infusion catheter was selected for use.The catheter was inserted into the patient, however, it was kinked so the guidewire was unable to pass through the catheter.A second direxion infusion catheter was then selected, however, when it was removed out of the hoop it was stretched.A third direxion was used to complete the procedure successfully.There were no patient complications reported and the patient was stable post procedure.However, device analysis revealed a fracture located 34.7 cm from the hub.
|
| |
|
Search Alerts/Recalls
|