MAUDE Adverse Event Report: CONCORD MANUFACTURING 2008K2 HEMODIALYSIS SYS OLC/DP, SPANISH; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONCORD MANUFACTURING 2008K2 HEMODIALYSIS SYS OLC/DP, SPANISH; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 190618
Device Problem Melted (1385)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/05/2019
Event Type  malfunction  
Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
 
Event Description
A biomedical technician (biomed) at a user facility reported that a fresenius 2008k2 hemodialysis (hd) machine shut off and had a burning smell during treatment while the patient was connected.The patient completed treatment on different machine.The machine was scheduled to service.Upon follow up, the machine was inspected and found to have melted fuses.There were no adverse events or medical intervention required as a result of the reported event.The biomed carried out functional tests and the machine performed as intended.
 
Manufacturer Narrative
Correction: preexisting medical conditions.
 
Manufacturer Narrative
Plant investigation: no parts were returned to the manufacturer for physical evaluation.A fresenius regional equipment specialist (res) mexico performed an onsite inspection, and found the machine had shorted fuses.The res resolved the issues by replacing the fuses.Testing and functional checks were performed.The machine functioned correctly and was returned to service.A serial number search in the ptc complaint system found no other complaints with the same symptoms within 90 days of the notified date.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2008K2 HEMODIALYSIS SYS OLC/DP, SPANISH
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
CONCORD MANUFACTURING
4040 nelson avenue
concord CA 94520
MDR Report Key8536271
MDR Text Key142641555
Report Number2937457-2019-01163
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
PMA/PMN Number
K994267
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup,Followup
Report Date 05/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/22/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number190618
Was Device Available for Evaluation? No
Device Age MO
Date Manufacturer Received05/09/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-