COVIDIEN MFG DC BOULDER COOL-TIP; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number ACT1520
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 04/03/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during ablation, the device had no power nor energy delivery.The same machine repeatedly turned on several times, showing the same problem, and after replacing a new set of electrode needles, it returns to normal.The procedure was aborted.
 
Additional Manufacturer Narrative
Evaluation summary: this report is based on information provided by post market vigilance investigation personnel.One device was received for evaluation.The returned product met specification as received.Visual inspection found no defects.The reported condition was not confirmed.The investigation found the water circulated normally through the sample.The device read the temperature and lesioned properly.The investigation found the device to function normally and within specifications.This unit does not meet the requirements for a manufacturing failure therefore a device history review is not required.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOL-TIP
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
MDR Report Key8537045
Report Number1717344-2019-00543
Device Sequence Number16751819
Product Code GEI
UDI-Device Identifier10884524001180
UDI-Public10884524001180
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K973297
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/06/2021
Device Model NumberACT1520
Device Catalogue NumberACT1520
Device Lot Number82840149X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/30/2019
Initial Date Received by Manufacturer 04/03/2019
Supplement Date Received by Manufacturer07/02/2019
Initial Report FDA Received Date04/22/2019
Supplement Report FDA Received Date07/18/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age62 YR
Patient Weight65
-
-