Model Number 364484
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/25/2019
Type of Reportable Event Malfunction
Event or Problem Description
Ous mdr - an orsiro drug-eluting stent system was selected for use.As it was difficult to connect the orsiro stent system with an indeflator the stent was visibly checked and a crack at the hub was noticed.
 
Additional Manufacturer Narrative
The device was not returned to biotronik.Therefore no technical investigation on the subject could be performed.The manufacturing history was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause was determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORSIRO 3.0/15
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key8537291
Report Number1028232-2019-01554
Device Sequence Number16576929
Product Code NIQ
UDI-Device Identifier07640130438996
UDI-Public07640130438996
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P170030
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/05/2020
Device Model Number364484
Device Catalogue NumberSEE MODEL NO.
Device Lot Number01182891
Initial Date Received by Manufacturer 04/15/2019
Supplement Date Received by Manufacturer05/20/2019
Initial Report FDA Received Date04/22/2019
Supplement Report FDA Received Date05/22/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-