| Model Number |
364484 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/25/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Ous mdr - an orsiro drug-eluting stent system was selected for use.As it was difficult to connect the orsiro stent system with an indeflator the stent was visibly checked and a crack at the hub was noticed.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned to biotronik.Therefore no technical investigation on the subject could be performed.The manufacturing history was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause was determined.
|
| |
|