| Model Number |
HEALON 0.55 |
| Medical Device Problem Code |
Separation Problem (4043)
|
| Health Effect - Clinical Code |
Capsular Bag Tear (2639)
|
| Date of Event |
02/19/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If implanted, give date: not applicable as this is not an implantable device.If explanted, give date: not applicable as this is not an implantable device.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that when the surgeon engaged, pulled out the healon plunger during retraction, the healon cannula shot off into the patient's eye and the force led to tear in the capsule.Reportedly, this altered the surgical plan of relaxing cut 90° away, iol sulcus optic capsule healon to maintain during changes and miostat was used.It was indicated that the surgery completed successfully and that the patient outcome is within the normal limits.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the product testing could not be performed as the device was not returned for evaluation.The reported complaint could not be verified.Manufacturing records review: the manufacturing records for the device were reviewed and no deviation was found.The product was manufactured and released according to specification.A search in complaint system revealed that no similar complaints have been received for this batch.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and (b)(4).
|
| |
|