| Catalog Number |
140.686 |
| Medical Device Problem Code |
Loss of Osseointegration (2408)
|
| Health Effect - Clinical Codes |
Pain (1994); Tissue Damage (2104)
|
| Date of Event |
04/01/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Missing lot information was requested from the initial reporter.A follow up report will be submitted when the information becomes available.Exemption number: e2015015.Instradent usa, inc is submitting the report on behalf of neodent - jjgc.
|
| |
|
Event or Problem Description
|
|
The clinician reported that 11 months after the dental implant was placed in the patient's mouth, the implant lost osseointegration in type ii bone.The clinician noted the patient's pain.The site was grafted.The product will be forwarded to the manufacturer for investigation.There were no reported post-operative complications.
|
| |
|
Additional Manufacturer Narrative
|
|
Exemption number: e2015015.Instradent usa, inc is submitting the report on behalf of neodent - jjgc.
|
| |
|
Event or Problem Description
|
|
The clinician reported that 11 months after the dental implant was placed in ada site 10 in the patient's mouth, the implant lost osseointegration in type ii bone.The clinician noted the patient's pain.The site was grafted.The product will be forwarded to the manufacturer for investigation.There were no reported post-operative complications.
|
| |
|
Additional Manufacturer Narrative
|
|
Exemption number: e2015015.Instradent usa, inc is submitting the report on behalf of neodent - jjgc.Considering the surgical methodology, it was seen that the dentist has selected an implant design which is not recommended for the patient's bone quality.The information provided in this report was based on information given by the dentist in neodent¿s products warranty form that was recorded in the system that is used by both the manufacturer and the subsidiary.The original complaint and the product were received by the subsidiary and did not arrive in the manufacturer yet.When this happens, a new evaluation of the complaint will be carried out and, if necessary, a new follow-up report will be send.
|
| |
|
Event or Problem Description
|
|
The clinician reported that 11 months after the dental implant was placed in ada site 10 in the patient's mouth, the implant lost osseointegration in type ii bone.The clinician noted the patient's pain.The site was grafted.The product will be forwarded to the manufacturer for investigation.There were no reported post-operative complications.
|
| |
|
Search Alerts/Recalls
|