• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MASIMO - 40 PARKER O3 REGIONAL OXIMETER MODULE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MASIMO - 40 PARKER O3 REGIONAL OXIMETER MODULE Back to Search Results
Model Number 24667
Medical Device Problem Code Thermal Decomposition of Device (1071)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/01/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product has been returned to masimo for evaluation.When the investigation is complete a follow up report will be submitted.
 
Event or Problem Description
It was reported that "o3 modules found with heat damage to the sensor connections." no consequences or impact to patient were reported.
 
Additional Manufacturer Narrative
The returned module was evaluated.Visual inspection found signs of damage that is consistent with the module subjected to drops and sensors connected backwards.The module was connected to a root and successfully established communication.  during testing the module was able to obtain measurements continuously with a known good sensor.The customer complaint not duplicated.The module is fully functional.  a service history record review reveals that this unit was in the field for over one (1) year with no previous reported issues related to this reported event.
 
Event or Problem Description
It was reported that "o3 modules found with heat damage to the sensor connections." no consequences or impact to patient were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O3 REGIONAL OXIMETER MODULE
Common Device Name
OXIMETER
Manufacturer (Section D)
MASIMO - 40 PARKER
40 parker
irvine CA 92618 1604
MDR Report Key8539885
Report Number2031172-2019-00213
Device Sequence Number10852895
Product Code MUD
UDI-Device Identifier00843997010771
UDI-Public00843997010771
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K160526
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other,use
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 04/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number24667
Device Catalogue Number9637
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/03/2019
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 04/01/2019
Supplement Date Received by Manufacturer05/08/2019
Initial Report FDA Received Date04/23/2019
Supplement Report FDA Received Date06/03/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-