MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065751763
Device Problems Obstruction of Flow (2423); Device Contamination with Chemical or Other Material (2944)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/20/2019
Event Type  malfunction  
Manufacturer Narrative
The sample has been received and in-house evaluation is in progress.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A customer reported that during surgery, the i/a handpiece was clogged.Foreign material, "ball", was found in the ia tip.The case was completed without patient harm.
 
Manufacturer Narrative
The foreign material was received and a visual assessment of the returned sample showed a small foreign material (smaller than 1x1mm) secured under a tape.The sample was sent for material analysis; conclusion pending.The i/a tip lot number (l/n) was not provided and could not be determined based on the information provided.Therefore, manufacturing information could not be obtained.The root cause cannot be determined conclusively.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
A round opaque particle was approximately 260um in diameter and was found to be 54.24% match with bac (benzalkonium chloride).Further evaluation identified it as mostly carbon and chlorine.However, the exact identification of the particle is unknown.The root cause cannot be determined conclusively.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key8540374
MDR Text Key143025807
Report Number2028159-2019-00731
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
PMA/PMN Number
K121555
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 09/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065751763
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/04/2019
Initial Date Manufacturer Received 03/25/2019
Initial Date FDA Received04/23/2019
Supplement Dates Manufacturer Received08/02/2019
09/04/2019
Supplement Dates FDA Received08/06/2019
09/09/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
CEN FMS PACK,ACT,.9U 45 BAL; CENTURION OZIL HP
Patient Age57 YR
-
-