Catalog Number 8065751763 |
Device Problems
Obstruction of Flow (2423); Device Contamination with Chemical or Other Material (2944)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 03/20/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
The sample has been received and in-house evaluation is in progress.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
|
|
Event Description
|
A customer reported that during surgery, the i/a handpiece was clogged.Foreign material, "ball", was found in the ia tip.The case was completed without patient harm.
|
|
Manufacturer Narrative
|
The foreign material was received and a visual assessment of the returned sample showed a small foreign material (smaller than 1x1mm) secured under a tape.The sample was sent for material analysis; conclusion pending.The i/a tip lot number (l/n) was not provided and could not be determined based on the information provided.Therefore, manufacturing information could not be obtained.The root cause cannot be determined conclusively.The manufacturer internal reference number is: (b)(4).
|
|
Manufacturer Narrative
|
A round opaque particle was approximately 260um in diameter and was found to be 54.24% match with bac (benzalkonium chloride).Further evaluation identified it as mostly carbon and chlorine.However, the exact identification of the particle is unknown.The root cause cannot be determined conclusively.The manufacturer internal reference number is: (b)(4).
|
|