• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONVATEC DOMINICAN REPUBLIC INC ACTIVELIFE; POUCH, COLOSTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONVATEC DOMINICAN REPUBLIC INC ACTIVELIFE; POUCH, COLOSTOMY Back to Search Results
Model Number 175779
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Based on the available information, this event is deemed to be a reportable malfunction.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.(b)(4).
 
Event or Problem Description
The end user reported an issue with five out of five precut ostomy appliances, where the opening was cut off center.
 
Additional Manufacturer Narrative
Additional information - this emdr is being submitted to include the below: d4: unique identifier (udi) h6: investigation results under imdrf cause investigation code, imdrf investigation findings, imdrf cause conclusions h10: investigation summary batch record review lot 8c02063 was manufactured on (b)(6) 2018 in the convex 1pc line with a total of 4,000 market units.Complaint investigator performed a batch record review on (b)(6) 2019 to verify if all the applicable procedures were followed and no issues were found; all the components for assembly were correct per bill of materials (bom) and all the tooling information documented was also correct, under international commodity code (icc) (b)(4) system application product (sap) material id (b)(4) and manufacturing order (b)(4).The batch record review supports that there were no discrepancies related to the issue reported.Returned sample evaluation: no photograph associated with this case and no unused return sample was expected.Conclusion summary of the related event: based on the investigation findings, the root cause identified for the issue ¿wafer off center in the pouch¿, reported under failure mode ¿skin barrier starter hole is defective (e.G misalignment or off center), leakage may occur¿ is attributed to: 1) machine: the investigative process concluded that all the machinery/ tooling items used in the manufacturing process complied when compared against drawing and process instruction (pi) pi21-076 rev 37.0 specifications; however, as part of this investigation on 10/may/2019, maintenance and facilities manager and senior process engineer performed a deep assessment to the sub-assembly station of the convex 1pc machine identifying the conditions causing the wafer off center and the possible solutions to correct the failure.As conclusion of the assessment, the chain that coordinates flange loading station with adhesive loading station needed lubrication and the chain and other components of the mechanism needed a corrective maintenance.In order to maintain this condition in the long term, a deeper intervention of the mechanism is required.Furthermore, regarding the rotary table station of the convex 1pc machine, it was identified that due to the demands of the process, the tooling (cup pouch uro 177a) used to manufacture the urodress product family require a redesign modification to reduce the variability within the manufacturing process.As a contributor factors the following opportunities for improvements were identified: 2) process / method investigation: a) reference in the pi21-076; section 8.0 ¿quality instructions¿ the applicable qc tooling used for product¿s quality inspection purpose of pouch uro 177a products family in order to address manufacturing personnel the use of them.B) identification of the defect during the manufacturing process, it should be considered the implementation of a job aid for product centralization out of specification.3) process / method / measurement investigation a) introduce qc tooling used for product¿s quality inspection purpose in the calibration program to guarantee measurement assurance.4) manpower: a) there are not manpower causes associated with the root cause; even though, opportunity to improve the loading of the adhesive disc during the manufacturing process was identified by implementing a standard work instruction for convex 1pc to indicate manufacturing operators how to properly load the adhesive disc.No issues were identified for material and environment investigations.The investigation associated with related event has been approved and is complete.No additional action is required, and this complaint will be closed.This issue will be monitored through the post market product monitoring review process.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number reporting site: (b)(4) manufacturing site: (b)(4).
 
Event or Problem Description
To date no additional patient or event details have been received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVELIFE
Common Device Name
POUCH, COLOSTOMY
Manufacturer (Section D)
CONVATEC DOMINICAN REPUBLIC INC
km 18.5 parque industrial
itabo, s.a. haina
san cristobal
Manufacturer (Section G)
CONVATEC DOMINICAN REPUBLIC INC
km 18.5 parque industrial
itabo, s.a. haina
san cristobal
Manufacturer Contact
pamela meadows
7815 national service road
suite 600
greensboro, NC 27409
3365424679
MDR Report Key8540379
Report Number9618003-2019-02887
Device Sequence Number367755
Product Code EZQ
UDI-Device Identifier00768455181254
UDI-Public00768455181254
Combination Product (Y/N)N
Initial Reporter StateNC
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date03/22/2023
Device Model Number175779
Device Lot Number8C02063
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/08/2019
Supplement Date Received by Manufacturer08/04/2023
Initial Report FDA Received Date04/23/2019
Supplement Report FDA Received Date08/24/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Patient SexFemale
-
-