| Model Number |
366133 |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/15/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The technical investigation of the returned device revealed that the balloon shows no signs of inflation.The device hub was inspected and found to be liquid tight.An inflation device was connected and while applying pressure a fine jet of water was seen to emerge from the center of the balloon.Microscopic analysis showed a pinhole in the balloon center and scratches on the balloon surface nearby the damage site.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.In addition to visual inspections each instrument is tested for air tightness by means of a helium leak test.Based on the conducted investigations of the device being subject to this complaint, no manufacturing or material related root cause was determined.The final root cause for the reported event is most likely related to the patients anatomy.
|
| |
|
Event or Problem Description
|
|
Ous mdr - as it was attempted to place the passeo-18 balloon catheter inside the lesion it was noticed that fluid was exiting from the hub.
|
| |
|