| Catalog Number |
H938099EG |
| Medical Device Problem Code |
Device Emits Odor (1425)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The actual device was returned for evaluation.A visual inspection was performed and no issues were noted.The device was disassembled, and it was noted that there were no signs of burns or fire inside of the pump and there was also no electrical smell.The reported condition was not verified.A device history review revealed no issues that could have caused or contributed to the reported issue.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
This report summarizes 1 malfunction events.It was reported that the repeater pump hardware had an electrical smell at high speed.There was no patient involvement.No additional information is available.
|
| |
|
Search Alerts/Recalls
|