Medtronic received report that the distal tip of the pipeline device poorly opened, the device was unable to be resheathed.During these maneuvers, the pipeline and the catheter fell into the aneurysm sac.The system was withdrawn from the body and the catheter was found to be deformed due to the patient's blood vessel being distorted (tortuous).The catheter¿s distal section was noted to be damaged when assessed.A new catheter and stent were used to complete the surgery.The anatomy was severe in tortuosity.The reported device and accessory devices were prepared as indicated in the ifu.The catheter was flushed as indicated in the ifu.More than 50% of the device was deployed when it failed to open.The device was resheathed less than or equal to 2 times.
|
|
D10: device available for evaluation - additional information.G4.Date mfr rec ¿ additional information.H2: type of follow up - device evaluation.H3: device evaluated by manufacturer- additional information.H6: codes updated the device was returned for evaluation and the clinical observation could not be confirmed.As received, the distal and proximal ends of the pipeline flex braid fully opened and moderately frayed.The pushwire appeared to be bent at multiple locations from the proximal end.The dps sleeves were found intact with no signs of damage.The outer diameter (o.D) of the re-sheathing pad was measured within specifications.The distal hypotube appeared to be stretched with the ptfe shrink tubing still intact.No damages were found with tip coil, the distal marker, re-sheathing marker or with the proximal bumper.The distal and proximal dps restraints were found to be intact.We were unable to determine the cause of pipeline flex device of failure to open during the procedure as returned pipeline braid fully opened and moderately frayed.The damage on the ends of the braid is likely the results of the physician re-sheathing the device more than recommended two times.It is possible that the severe vessel tortuosity may have contributed to the failure to open and resheathing issues.Per instructions for use (ifu), ¿begin to deliver the device using a combination of unsheathing the pipeline flex embolization device and pushing delivery wire simultaneously.After the distal end of the pipeline flex embolization device has successfully expanded, deploy the remainder of pipeline flex embolization device by pushing the delivery wire and/or unsheathing the pipeline flex embolization device.Re-sheathing and/or manipulation of the micro catheter, by locking down the delivery wire and moving both as a system, may facilitate expansion of the pipeline flex embolization device.Carefully inspect the deployed pipeline flex embolization device under fluoroscopy to confirm that it is completely apposed to the vessel wall and not kinked.If the device is n ot fully apposed or is kinked, consider using a balloon catheter, micro catheter, or guide-wire to fully open it.Do not use in patients in whom the angiography demonstrates the anatomy is not appropriate for endovascular treatment, due to conditions such as severe intracranial vessel tortuosity or stenosis.¿ medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|