| Model Number |
MMT-523NAS |
| Medical Device Problem Codes |
Premature Discharge of Battery (1057); No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/16/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
Customer reported via phone call that the insulin pump had a blank display.The customer¿s blood glucose level was unknown at the time of the incident.The customer was assisted with troubleshooting and customer reports display was blank currently.Customer states the display was not blank less than 30 seconds and did not return.The customer removes the battery and stated the insert battery alarm did not display on the insulin pump screen.Customer states the contact on the battery cap was neither missing nor damage.Customer states battery compartment was neither damaged nor corroded.Customer states the spring was neither damaged nor corroded.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The customer was advised the insulin pump will be replaced as per troubleshoot.The insulin pump will be returned for analysis.
|
| |
|