Catalog Number 136561000
Medical Device Problem Codes Naturally Worn (2988); Noise, Audible (3273)
Health Effect - Clinical Codes Cyst(s) (1800); Pain (1994); Tissue Damage (2104); Weakness (2145); Hypoesthesia (2352); Limited Mobility Of The Implanted Joint (2671); No Code Available (3191)
Date of Event 09/18/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Initial reporter occupation: lawyer.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Claim and medical records received.After review of claim and medical records, it was stated that the patient was revised to address pain and aseptic loosening of cup component.Stiffness, limping, clicking, popping, weakness and numbness were also noted.Operative notes reported slightly brown tented and translucent fluid, the tissue is stained dark gray consistent of metallosis, two superior cysts were curetted and there was no acetabular bony ingrowth noted.When removing the acetabular shell, three screws broke but only one was fully extracted and the two screws were recessed since they were embedded into the sclerotic bone.Mri reported trochanteric bursitis.X-ray reported aseptic loosening of acetabular component and has metal wear debris reaction.Doi: (b)(6) 2010; dor: (b)(6) 2018; left hip.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
12/14 ARTICUL 44MM M SPEC+1.5
Common Device Name
ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key8544307
Report Number1818910-2019-91467
Device Sequence Number400745
Product Code JDI
UDI-Device Identifier10603295034032
UDI-Public10603295034032
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K060031
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2018
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2012
Device Catalogue Number136561000
Device Lot Number2360843
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/02/2019
Supplement Date Received by Manufacturer05/09/2019
05/30/2019
Initial Report FDA Received Date04/24/2019
Supplement Report FDA Received Date05/09/2019
05/30/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age76 YR
Patient Weight98
-
-