• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY FRANCE SAS - 3003895575 CORAIL REVISION STEM HO 13; CORAIL KAR REVISION IMPLANT : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY FRANCE SAS - 3003895575 CORAIL REVISION STEM HO 13; CORAIL KAR REVISION IMPLANT : HIP FEMORAL STEM Back to Search Results
Catalog Number L98113
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 04/03/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Fractured corail revision stem.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: visual examination of the returned stem confirms the reported material fracture.A metallurgical evaluation found no material or manufacturing defects that could have contributed to fracture initiation and/or propagation to failure.The device manufacturing records and material certification have been reviewed.No related deviations or anomalies were identified.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORAIL REVISION STEM HO 13
Common Device Name
CORAIL KAR REVISION IMPLANT : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY FRANCE SAS - 3003895575
7 allee irene joliot-curie
b.p. 256
saint priest cedex 69801
FR  69801
MDR Report Key8545246
Report Number1818910-2019-91504
Device Sequence Number367986
Product Code LWJ
UDI-Device Identifier10603295262411
UDI-Public10603295262411
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K093736
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberL98113
Device Lot Number8072842
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/26/2019
Initial Date Received by Manufacturer 04/03/2019
Supplement Date Received by Manufacturer04/26/2019
06/24/2019
Initial Report FDA Received Date04/24/2019
Supplement Report FDA Received Date04/29/2019
06/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-