Device evaluated by mfr.: returned product consisted of an emerge balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed multiple kinks along the hypotube.The hypotube is separated 64.5cm from the hub and appears to have been kinked prior to separation.Microscopic examination revealed no additional damages.There is contrast present in the inflation lumen and on the balloon folds.There is blood in the guide wire lumen and the balloon is still tightly folded.Inspection of the remainder of the device presented no other damage or irregularities.
|
Reportable based on device analysis completed on (b)(6) 2019.It was reported that balloon damage occurred.Vascular access was obtained via the right radial artery.The target lesion was located in the moderately tortuous and moderately calcified left circumflex artery.A 1.50mm x 15mm emerge balloon catheter was advanced for dilatation.However, during retreat of the device form the vessel, the balloon catheter got damaged.The procedure was completed with another of the same device.No patient complications were reported.However, returned device analysis revealed shaft detached/separated.
|