• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-KALAMAZOO DISP. 30X4, 1BLA, 1PRT, QUICK CONNECT; TOURNIQUET, PNEUMATIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-KALAMAZOO DISP. 30X4, 1BLA, 1PRT, QUICK CONNECT; TOURNIQUET, PNEUMATIC Back to Search Results
Catalog Number 5921030135
Medical Device Problem Code Decrease in Pressure (1490)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/01/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Reported events: 1 event was reported for this quarter.Product return status: 1 device was received.Event confirmation status: 1 reported event was confirmed; the cause traced to component failure.Evaluation results: 1 device was found to be affected by a mechanical problem.  additional information: 1 device is labeled for single-use.1 device was not reprocessed and reused.
 
Event or Problem Description
This report summarizes 1 malfunction event in which the device decreased in pressure.1 event had patient involvement; no patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISP. 30X4, 1BLA, 1PRT, QUICK CONNECT
Common Device Name
TOURNIQUET, PNEUMATIC
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
MDR Report Key8545859
Report Number0001811755-2019-01344
Device Sequence Number1744156
Product Code KCY
UDI-Device Identifier37613154599418
UDI-Public37613154599418
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)Y
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Catalogue Number5921030135
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/01/2019
Initial Report FDA Received Date04/24/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-