| Catalog Number |
5400210056 |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
01/01/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Reported events: 4 events were reported for this quarter.Product return status: 4 devices were received.Event confirmation status: 4 device evaluations are still in progress.Additional information: 4 devices were not labeled for single-use.4 devices were not reprocessed and reused.
|
| |
|
Event or Problem Description
|
|
This report summarizes 4 malfunction events in which the device reportedly overheated.4 events had no patient involvement; no patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Supplemental rationale: 4 previously reported events are included in this follow-up record.Product return status: 4 devices were received.Event confirmation status: 4 reported events were not confirmed.Evaluation results: 3 devices were found to be affected by internal component corrosion.1 device was found to be affected by internal debris.
|
| |
|
Event or Problem Description
|
|
This report summarizes 4 malfunction events in which the device reportedly overheated.4 events had no patient involvement; no patient impact.
|
| |
|