| Catalog Number |
5400218000 |
| Medical Device Problem Code |
Device Remains Activated (1525)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
01/01/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Reported events: 1 event was reported for this quarter.Product return status: 1 device was received.Event confirmation status: 1 device evaluation is still in progress.Additional information: 1 device is not labeled for single-use.1 device was not reprocessed and reused.
|
| |
|
Event or Problem Description
|
|
This report summarizes 1 malfunction event in which the device had run-on.1 event had no patient involvement; no patient impact.
|
| |
|
Event or Problem Description
|
|
This report summarizes 1 malfunction event in which the device had run-on.1 event had no patient involvement; no patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Supplemental rationale corrected data: 1 previously reported event is included in this follow-up record.Product return status 1 device was received.Event confirmation status 1 reported event was confirmed.Evaluation results 1 device was found to be affected by internal components affected by debris.
|
| |
|
Search Alerts/Recalls
|