| Catalog Number |
7203126000 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
01/01/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Reported events: 1 event was reported for this quarter.Product return status: 1 device was received.Event confirmation status: 1 reported event was not confirmed.Evaluation results: 1 device had no device problem found.Additional information: 1 device is not labeled for single-use.1 device was not reprocessed and reused.
|
| |
|
Event or Problem Description
|
|
This report summarizes 1 malfunction event in which the device was reportedly leaking.1 event had no patient involvement; no patient impact.
|
| |
|
Search Alerts/Recalls
|