• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONCORD MANUFACTURING LIBERTY SELECT CYCLER ASSY(NON-VALUATED); SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONCORD MANUFACTURING LIBERTY SELECT CYCLER ASSY(NON-VALUATED); SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number RTLR180343
Device Problem Overfill (2404)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/15/2019
Event Type  malfunction  
Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Event Description
This is a report of a patient who experienced an increased intraperitoneal volume (iipv) event coincident with peritoneal dialysis (pd) therapy.The event was discovered while reviewing the patient¿s treatment records following a report of a patient having a patient line is blocked alarm during drain 3 of 4.The patient unintentionally bypassed the alarm and experienced a large drain during drain 4.A review of the patient¿s treatment records identified that the patient drained 4287 ml during drain 4 of the treatment.This drain volume is 214% the patient's prescribed fill volume of 2000 ml.As a result of the iipv event, the technical support representative advised the patient to discontinue use of the cycler and to notify their peritoneal dialysis registered nurse (pdrn) of the event.Upon follow up with the pdrn, it was confirmed that the patient did not experience any symptoms, adverse events, injuries, or require medical intervention as a result of the reported event.The pdrn stated that the patient is trained on performing stat drains, as well as manual peritoneal dialysis therapy if needed, and stated the patient was able to complete on the cycler.The patient has received a new cycler which is working well and continuing with pd therapy without issue.The cycler was reported to be returned to the manufacturer for physical evaluation.
 
Manufacturer Narrative
The actual device was returned to the manufacturer for physical evaluation.An external visual inspection was performed with no physical damage noted.A simulated therapy was initiated and completed on the cycler without complication.The weighed fill volumes were found to be within tolerance and fill times were within specification.The cycler underwent system air leak and valve actuation testing and was found to meet product specifications.Load cell verification testing was performed with no issues noted.A review of the device manufacturing records was conducted by the manufacturer.There were no deviations or non-conformances during the manufacturing process.In addition, a device history record (dhr) review was performed and verified that the results of the in-progress and final quality control (qc) testing met all requirements.Upon completion of the evaluation, there were no malfunctions that could have caused or contributed to the reported event.The cycler performed as designed and an associated cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIBERTY SELECT CYCLER ASSY(NON-VALUATED)
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
CONCORD MANUFACTURING
4040 nelson avenue
concord CA 94520
Manufacturer (Section G)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer Contact
matthew amaral
920 winter st.
waltham, MA 02451
7816999758
MDR Report Key8546507
MDR Text Key143603856
Report Number2937457-2019-01191
Device Sequence Number1
Product Code FKX
UDI-Device Identifier00840861102068
UDI-Public00840861102068
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K181108
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 05/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/24/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberRTLR180343
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/24/2019
Device Age MO
Date Manufacturer Received05/04/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/04/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age71 YR
Patient Weight99
-
-