| Model Number |
4FC12 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Death (1802); Cardiac Perforation (2513)
|
| Date of Event |
04/11/2019
|
|
Type of Reportable Event
|
Death
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that during a cryo ablation procedure, the right inferior pulmonary vein (ripv) was perforated.Additionally, the physician noticed blood in the endotracheal (et) tube.Blood was suctioned from the et tube, but bleeding continued, and the case had to be aborted while the patient was under general anesthesia.The patient is deceased.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the data files were returned and analyzed.The data files showed eight applications were performed with balloon catheter 2af284 with lot number 96020 without any issues present.A clinical issue (cardiac perforation and death) was encountered during the procedure.The case was aborted while the patient was under general anesthesia.In conclusion, the reported sheath was not returned for investigation.There is no indication of relation of the adverse event to the performance of the cryo device.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|