| Model Number |
MT22495 |
| Medical Device Problem Code |
Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/03/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the receiver display was frozen.No product or data was provided for evaluation.Confirmation of the allegation and probable cause could not be determined.No injury or medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the receiver ceased to function.The product was evaluated.An external visual inspection was performed and passed.A receiver charge and boot was performed and passed.A functional test was performed and passed relevant testing.A receiver download was performed and passed finding no errors related to the complaint.Confirmation of the complaint was not confirmed.The probable cause could not be determined.
|
| |
|