AVANOS MEDICAL - IRVINE ON-Q PAIN RELIEF SYSTEM WITH SELECT-A-FLOW, 400 ML, 2-14 ML/HR; ELASTOMERIC - SAF Back to Search Results
Model Number CB004
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The actual complaint product was not returned for evaluation.A review of the device history record is not possible as no lot number was provided.Root cause could not be determined.All information reasonably known as of 24-apr-2019 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).
 
Event or Problem Description
Avanos medical, inc.Received a single report that referenced three different incidences, which were associated with separate units, involving three different events.This is the third of three reports.Refer to 2026095-2019-00064 for the first event.Refer to 2026095-2019-00065 for the second event.Fill volume: 550 ml, flow rate: 6 ml/hr, procedure: unknown, cathplace: unknown.It was reported the onq pumps were running out too quickly.The pumps were completely infusing on post-operative day (pod) three instead of on pod four.Additional information received 29-mar-2019 stated the onq pumps appeared to be running out earlier than expected given the set rate, only lasting until pod#3 at rate of 6 ml/hr on 550ml fill and should have lasted until pod#4.There was no injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ON-Q PAIN RELIEF SYSTEM WITH SELECT-A-FLOW, 400 ML, 2-14 ML/HR
Common Device Name
ELASTOMERIC - SAF
Manufacturer (Section D)
AVANOS MEDICAL - IRVINE
43 discovery
suite 100
irvine CA 92618
MDR Report Key8547596
Report Number2026095-2019-00066
Device Sequence Number1731151
Product Code MEB
UDI-Device Identifier30680651134722
UDI-Public30680651134722
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K063530
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCB004
Device Catalogue Number101347200
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/01/2019
Initial Report FDA Received Date04/24/2019
Patient Sequence Number1
-
-