This follow-up report is being filled to relay investigation result.Dhr review: the dhr check could not be performed as the lot number was not available.Trend analysis: no trend considering the following event is identified: jammed.No trend considering the following event is identified: fracture.Event description: it was reported that during a revision surgery on (b)(6) 2019 when surgeon tried to remove the stem trial after finishing trial-reduction, the screw broke inside the distal trial, and made the distal trial cannot be retrieved from the canal.Surgeon had no choice to do osteotomy to open the bone to take out the instrument.There was no surgical delay in excess of 30 minutes.Review of received data: no medical data provided due to country regulations.Device analysis: visual examination: the wagner sl trial stem d16 shows some major scratches and missing parts of the four flanks.Further, the broken screw inside the trial stem can be seen.Based on this visual examination the reported event can be confirmed.Measurements: wagner sl trial stem d16: to ensure the wagner sl trial stem d16 has correct dimensions, relevant characteristics according to inspection plan sap document no.4973 rev.04 and product drawing sap doc.No.063.223.184 rev 04 were measured with caliper so 2083.Characteristic no.3 feature innen ø at cross-section c-c specification: max.16.2 mm; min.15.9 mm.Measured value: 16.13 mm.Conclusion: the characteristic of the wagner sl trial stem d16 is within specification.Screw: no measurement was performed on the screw, as the screw is stuck inside the trial stem and the broken off part of the screw has not been returned.Functional test with the same device showed, that when the screw is not screwed in completely, thus incorrect instrument handling, there is clearance between the distal and the proximal part, which resulted in higher loads leading to screw breakage.Review of product documentation: all involved devices are intended for treatment.Compatibility: compatibility check could not be performed as the product identification is given for the distal trial stem only.The reported screw for proximal trial stem l225 (ref: 01.00109.805) must be used with the proximal trial stem of length 225 (ref: 01.00109.802).As the product identifications of the screw as well as of the proximal trial stem are unknown or could not be reviewed, we cannot confirm the compatibility of the used instruments.Inspection plan sap: characteristic no.3 feature innen ø with scope of testing: 100%.Means of inspection: caliper.Ordering information sap: the screws for proximal trial stems are size dependent.For instance, the screw for the proximal trial stem l225 (ref: 01.00109.805) must be used with the proximal trial stem of length 225 (ref: 01.00109.802) only.Root cause analysis: root cause determination using sap rmw: instrument breaks, deforms, diverges impairing its function.Due to inadequate design for intended performance = not possible, as a systematic issue with design would have been detected as part of the issue evaluation assessment defined in complaint investigation or in the current pms process.Instrument, breaks and parts remain in wound.Due to inadequate design for intended performance = not possible, as according to material compatibility specification the material has been tested.Further, a systematic issue with material properties would have been detected as part of the complaint investigation or in the current pms process.Instrument breaks, deforms, diverges impairing its function due to mechanical properties of material insufficient = not possible, as according to material compatibility specification the material has been tested.Further, a systematic issue with material properties would have been detected as part of the complaint investigation or in the current pms process.Instrument, breaks and parts remain in wound due to mechanical properties of material insufficient = not possible, as according to material compatibility specification , the material has been tested.Further, a systematic issue with material properties would have been detected as part of the complaint investigation or in the current pms process.Fracture of instrument due to general corrosion (crevice, pitting, galvanic) = not possible, there were no signs of corrosion found during investigation.Damaged instruments, implants, body or wrong operational step due to surgeon or or staff unfamiliar with instrument usage and handling = possible, based on the given information this cause cannot be excluded.Damaged instruments, implants, body or wrong operational step due to surgeon or or staff unfamiliar with instrument usage and handling = possible, based on the given information this cause cannot be excluded.Instrument breaks or deforms due to off-label / abnormal-use = possible, based on the given information this cause cannot be excluded.Components stuck together, incorrect conclusion on size due to compatibility of incompatible combinations = possible, as the screw can no longer be retrieved this cause cannot be excluded.Conclusion: based on the returned product and the given information the complaint could be confirmed.The visual examination did confirm that the screw is broken and stuck inside the distal trial stem.The damages found during the visual examination on the distal trial stem were introduced during removal of the stem upon osteotomy.The quality records of the distal trial stem show that all specified characteristics (material, dimensions, surface, etc.) have met the specifications valid at the time of production.The quality records of the screw and of the proximal trial stem could not be reviewed as the given device identifications are insufficient.As the proximal trial stem has neither been received nor reported and the device identification of the screw is unknown, it is possible that the wrong combination of screw and proximal trial stem size was used.Further, it may be possible that incorrect assembly of the instruments, which may lead to higher loads or bending stresses than accepted, caused the screw to break.In conclusion, based on the investigation and the missing device identification of the screw and the proximal trial stem, we were not able to identify a specific root cause for the reported event.The need for corrective measures is not indicated and zimmer gmbh considers this case as closed.Zimmer biomet¿s reference number of this file is (b)(4).
|