• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS PROLACTIN ASSAY; RADIOIMMUNOASSAY, PROLACTIN (LACTOGEN)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS PROLACTIN ASSAY; RADIOIMMUNOASSAY, PROLACTIN (LACTOGEN) Back to Search Results
Model Number PROLACTIN G2
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigations for the 1 event did not identify a product problem.The cause of the event could not be determined.The patient sample was requested for investigation but was not provided.The follow up/corrective actions for the 1 event were as follows: the patient sample was run on a competitor system.A manual dilution was performed and the sample was checked for hemolysis.Calibration and qc were acceptable.No instrument issues were detected.The results for other patients were normal.No devices were returned.This device is not labeled for single use and is not reprocessed or reused.
 
Event or Problem Description
This report summarizes 1 malfunction event.An erroneous low result was generated by the cobas 6000 e 601 module.The events involved a total of 1 patient with the following: a low prolactin result.The patient's age was (b)(6).The patient's gender was female.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS PROLACTIN ASSAY
Common Device Name
RADIOIMMUNOASSAY, PROLACTIN (LACTOGEN)
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250
MDR Report Key8549741
Report Number1823260-2019-90089
Device Sequence Number13849613
Product Code CFT
UDI-Device Identifier04015630921546
UDI-Public04015630921546
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)Y
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device No Information
Device Model NumberPROLACTIN G2
Device Lot Number30128002
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date04/25/2019
Patient Sequence Number1
-
-