Device issue with inomax dsir (b)(4).Was documented and investigation under (b)(4).A review of the device's service log revealed that a delivery failure alarm occurred due to over-delivery of nitric oxide (no) for 12 consecutive seconds.Although identified in the service log, the regional service center (rsc) did not experience a delivery failure alarm during testing.The root cause for this occurrence was likely due to a malfunctioning proportional valve.As a result, the device's proportional valve was replaced.This condition will be tracked and trended under the company's quality system.A full functional test was performed and the device operated according to specifications.As a result, the device was placed back into circulation for customer use.
|