U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MASIMO - 40 PARKER O3 REGIONAL OXIMETER MODULE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MASIMO - 40 PARKER O3 REGIONAL OXIMETER MODULE Back to Search Results
Model Number 24667
Medical Device Problem Code Incorrect Measurement (1383)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/26/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The returned module was evaluated.The module was connected to a root device and communication was successfully established.Both channels were able to take measurements which were consistent with measurements from a known good o3.After adding humidity to the connector side, the values were inconsistent with measurements from a known good o3.Internal inspection revealed that the cable connector had signs of contamination damages.The o3 module didn't work correctly and showed wrong values due to the contamination damages to the cable connector.Initial reporter phone number exceeded the maximum allowable characters, the phone number is as follows: (b)(6).
 
Event or Problem Description
The customer reported that the module didn't work.The module was returned for evaluation and it was determined that the module had contamination which resulted in inaccurate measurements.No consequences or impact to patient were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O3 REGIONAL OXIMETER MODULE
Common Device Name
OXIMETER
Manufacturer (Section D)
MASIMO - 40 PARKER
40 parker
irvine CA 92618 1604
MDR Report Key8553081
Report Number2031172-2019-00218
Device Sequence Number374014
Product Code MUD
UDI-Device Identifier00843997010771
UDI-Public00843997010771
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K160526
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 02/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number24667
Device Catalogue Number9637
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/05/2019
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/01/2019
Initial Report FDA Received Date04/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-