| Model Number |
24667 |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/26/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The returned module was evaluated.The module was connected to a root device and communication was successfully established.Both channels were able to take measurements which were consistent with measurements from a known good o3.After adding humidity to the connector side, the values were inconsistent with measurements from a known good o3.Internal inspection revealed that the cable connector had signs of contamination damages.The o3 module didn't work correctly and showed wrong values due to the contamination damages to the cable connector.Initial reporter phone number exceeded the maximum allowable characters, the phone number is as follows: (b)(6).
|
| |
|
Event or Problem Description
|
|
The customer reported that the module didn't work.The module was returned for evaluation and it was determined that the module had contamination which resulted in inaccurate measurements.No consequences or impact to patient were reported.
|
| |
|