| Model Number |
TRUBALANCE |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/01/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(manufacturer narrative = t, corrected data = f) internal report # (b)(4).Returned meter evaluated with no defect found.Test strips not returned for evaluation.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.Test strip udi #: (b)(4).Note: manufacturer contacted customer in a follow-up call to ensure that the replacement products resolved the initial concern - able to establish contact with customer who indicated that is comfortable with results from the replacement meter.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for high blood glucose test results.The customer is concerned with tests results from results obtained of 406 mg/dl.The customer's expected fasting blood glucose test result range is 150 - 211 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on a blood test was performed by the customer fasting and produced test result of 329 mg/dl.The customer declined to perform a second blood test.The product is stored according to specification in the bedroom.The test strip lot manufacturer's expiration date is 06/30/2020 and open vial date is 03/18/2019.The meter memory was reviewed for previous test result history: (b)(6).
|
| |
|