• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN EVERCROSS 035; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN EVERCROSS 035; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Catalog Number AB35W04080135
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/19/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the physician was using an evercross 4x80 balloon to treat a calcified 50% stenotic lesion in the mid popliteal artery.Tortuousity and calcification were described as severe.There was no damage noted to the packaging and no issues noted when removing the device from the tray.The device was prepped without issue.No embolic protection was used.A non-medtronic inflation device was used.The device did not pass through a previously deployed stent and no resistance was encountered, nor excessive force applied, during delivery of the device to the lesion.It was reported that the balloon burst at 6atm.The burst was described as a pin hole.All fragments of the balloon were retrieved.The procedure was completed using another 4x80 evercross.No patient injury was reported.
 
Additional Manufacturer Narrative
Additional information: the device was safely removed from the patient without intervention required.No fragmentation of the balloon occurred.The vessel was post-dilated.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERCROSS 035
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
MDR Report Key8554462
Report Number2183870-2019-00189
Device Sequence Number2169317
Product Code LIT
UDI-Device Identifier00821684061220
UDI-Public00821684061220
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K082579
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/10/2022
Device Catalogue NumberAB35W04080135
Device Lot NumberA781027
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/19/2019
Supplement Date Received by Manufacturer04/26/2019
Initial Report FDA Received Date04/26/2019
Supplement Report FDA Received Date05/21/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-